Compliance with strict regulatory guidelines and quality requirement demands accurate measurement, secure data, comprehensive reporting and alarms notification unlike any other product.
NECLOG provides variety of physical parameter to be logged, reported and alarm notification. Temperature, Humidity, Differential pressure, CO2 etc.
Neclog is one of the unique solution with many advantages as easy-to-use, cost effective, high-accuracy solution for mapping critical and FDA-regulated environments form stability chambers to freezers and warehouses.
Simple configuration and connection of multiple sensors on serial link makes it easy to use and hassle free.
On line data on base unit and on PC allows to track performance in real time.
NECLOG maintains all data and events in secure files, providing temperproof security with presentation ready reports to meet 21CFR Part 11.
High-speed data transfer
Significantly faster than serial(RS-485) communication
Global standard for communication protocol
No need for a separate PC interface
Data Operations from any PC over the LAN
Complete Operations driven through PLC
No Data Loss: Data Storage of upto 1000 readings on the PLC itself
At an interval of 30 mins, data upto 20 days is stored on the PLC itself
Facility to send emails directly from the PLC
All operations through a single 4" Color touch screen display
PLC IP Address setting
Component Working Hours
Standby system change over
Password protection for set values and calibration mode
When an alarm is triggered, an e-mail is sent to the registered PC or mobile address
*connection to a mail server is required to use e-mail alert.
Temperature & Humidity Deviations
Door Opening
Water Supply Failure
Component Failure
Mains Failure
Mobile Alert system
Time of Alarm transmission
Name of Equipment
Transmitted Message (Equipment Name, Alarm time, Type of Alarm & Limits)
Galary
Standard Equipment
Double walled with 80mm PUF insulated modular panels with easy on-site assembly | System is made up of split type condenser unit in well-ventilated area to avoid throw of hot air in the room |
SS 304 Interior & Exterior either SS 304 or Powder Coated G.I. | Stand by humidity system |
Stainless Steel Trays | Complete Operations driven through PLC with data logging capacity of upto 1000 readings & email facility |
Observation window - 300 x 300 mm | High speed Ethernet based communication |
Illuminated working area | All operations through a single 4" Color Touch Screen |
U-shaped SS tubular heaters | High-quality German Make temperature humidity sensor with 4-20mA output |
Flooring of Linoleum carpet | Safety devices: Low & High temperature cut-off settable thermostats & high humidity cut-off device |
Forced air circulation for uniform temperature | Audio/Visual alarms for deviations |
Non Condensing steam injection type humidity system with standby heater & auto switch overs | Complete with IQ, OQ, PQ & Operations & Maintenance Manual |
CFC free cooling system consisting of hermetically sealed compressor coupled with evaporation coil & condenser |
Optional Accessories
ICDAS - A Common Software suitable for 20 nos. Newtronic make equipments | Door Access system |
8pt. Temperature & Humidity Scanners | Mobile Alert system |
Stand by Cooling System with auto switch overs | Remote alarm system suitable for 4 nos. Newtronic make equipments |
Selection Guide
Models |
Capacity |
Internal Chamber |
External Chamber |
Tray Size |
Trays |
|
---|---|---|---|---|---|---|
Cuft. |
Litres |
HxWxD (mm) |
HxWxD (mm) |
(mm) |
Nos. |
|
NEPL 2190RSI | 190 | 5400 | 2000x1360x2000 | 2120x1480x2120 | 870x300 | 24 |
NEPL 2230RSI | 230 | 6400 | 2400x1360x2000 | 2520x1480x2120 | 870x300 | 32 |
NEPL 2280RSI | 280 | 8000 | 2000x2000x2000 | 2120x2120x2120 | 870x600 | 24 |
NEPL 2340RSI | 340 | 9600 | 2400x2000x2000 | 2520x2120x2120 | 870x600 | 32 |
NEPL 2428RSI | 428 | 12000 | 2400x2500x2000 | 2520x2620x2120 | 870x800 | 32 |
NEPL 2530RSI | 530 | 15000 | 2400x2500x2500 | 2520x2620x2620 | 725x800 | 48 |